Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, Recall
Recall Date
Apr 8, 2026
Company
Mentor Texas, LP.
Source
FDA
Reason: Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Reason for Recall
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated April 14, 2026
- CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc · ID FDA-DEV-Z-1684-2026
- Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, · ID FDA-DEV-Z-1685-2026
More recalls from Mentor Texas, LP.
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.