Medical Device Recalls
161 recalls found
TruSystem 7000 U Operating Table (dV), Product Code 1723633
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only,
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs
Products lack 510(k) clearance.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA
Kits contain infusion sets which were recalled for performance issues.
CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010
A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, ANTEPARTUM IV , CATALOG # MA1DH7533C; 2) KIT,
Kits contain infusion sets which were recalled for performance issues.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, ANESTHESIA, CATALOG # IVSVM41612; 2) KIT, ANESTHESIA, C
Kits contain infusion sets which were recalled for performance issues.
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) IV START LD KIT , CATALOG # IVS5342277; 2) IV
Kits contain infusion sets which were recalled for performance issues.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT, CABG, , CATALOG # PCOCCG3A12; 2) KIT, CABG,
Kits contain infusion sets which were recalled for performance issues.
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
EVOS 3.5MM LATERAL PROXIMAL TIBIA PLATE 6 HOLE LEFT 91MM REF72463106
Incorrect labeling; 3.5MM Partial Articular Lateral Proximal Tibia Plate 6 hole Left 91MM implant labeled and lasered etched as as 3.5MM Lateral Proximal Tibia Plate 6 hole left 91mm implant
Generic Name: Perma Sharp Sutures; Generic Name: PGA Absorbable Surgical Suture; Model/Catalog Number: PSN397V (Henry Schein Item No. 600-0331); Product Description: Sterile, synthetic, absorbable,
Products lack 510(k) clearance.
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8695P (HSI Item No. 600-9795); Product Description: Sterile, synthetic, nonabs
Products lack 510(k) clearance.
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT,C SECTION ADMIT,, CATALOG # MATJV8300D; 2) KIT,C SECTIO
Kits contain infusion sets which were recalled for performance issues.
Senographe Pristina
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8384P (HSI Item No. 600-7664); Product Description: Sterile, synthetic, nonabs
Products lack 510(k) clearance.
TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
Philips Azurion 7 B20. Model Numbers: 722068, 722226.
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
PENTAX Medical Ultrasound Endoscope. Model Numbers. EG-3870UTK, EG-3670URK.
Warnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.
ETHILON Nylon Suture, Product Code 697G
Nylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.