Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695 Recall
This is part of a larger recall. See all affected products.
Recall Date
Nov 11, 2026
Company
Baxter Healthcare Corporation
Source
FDA
Units
77 units
Reason: There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Reason for Recall
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 77 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated November 17, 2026
More recalls from Baxter Healthcare Corporation
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087 · Baxter Healthcare Corporation · Nov 11, 2026
- Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Descript · Baxter Healthcare Corporation · Oct 21, 2026
- Dextrose Injection, USP, 70 %, 2000 mL bags, Rx Only, Baxter Healthcare Corporation Deerfield, IL, 60016, Made in USA, NDC 00338-0719-06 · Baxter Healthcare Corporation · Sep 16, 2026
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.