BD Kiestra" ReadA; Catalog No.: 446948. Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Apr 8, 2026

Company

BD KIESTRA LAB AUTOMATION

Source

FDA

Units

82 units (7 US, 65 OUS)

Reason: In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.

Reason for Recall

In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.

Product: BD Kiestra" ReadA; Catalog No.: 446948.
Official Recall ID: FDA-DEV-Z-1658-2026
Units affected: 82 units (7 US, 65 OUS)

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 82 units (7 US, 65 OUS)

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated April 14, 2026