Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

May 6, 2026

Company

Hologic, Inc

Source

FDA

Units

1200 units

Reason: It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Reason for Recall

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Product: Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes
Official Recall ID: FDA-DEV-Z-1949-2026
Units affected: 1200 units

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 1200 units

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated May 12, 2026

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored

Stay Informed

Weekly recall digest — free, no spam.