LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

May 20, 2026

Company

MICROVENTION INC.

Source

FDA

Units

38

Reason: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

Reason for Recall

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

Product: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm
Official Recall ID: FDA-DEV-Z-2200-2026
Units affected: 38

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 38

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated May 25, 2026

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored