Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B Recall
Recall Date
May 20, 2026
Company
Medtronic Neuromodulation
Source
FDA
Units
10165 units
Reason: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
Reason for Recall
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 10165 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated May 25, 2026
More recalls from Medtronic Neuromodulation
- InterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 3560030 · Medtronic Neuromodulation · Oct 28, 2026
- Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor · Medtronic Neuromodulation · Jun 3, 2026
Similar recalls
- Philips Azurion 7 M12. Model Numbers: 722078, 722223. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M20. Model Numbers: 722228, 722281. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C · Canon Medical System, USA, INC. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V · Canon Medical System, USA, INC. · Nov 4, 2026
- CareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantable · Medtronic, Inc. · Nov 4, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.