Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Jun 10, 2026

Company

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Source

FDA

Units

27 units

Reason: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Reason for Recall

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

Product: Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Official Recall ID: FDA-DEV-Z-2285-2026
Units affected: 27 units

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 27 units

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated June 16, 2026

All affected products in this recall
  • Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2288-2026 · 51 units units
  • Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2284-2026 · 16 units units
  • Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2285-2026 · 27 units units

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored

Stay Informed

Weekly recall digest — free, no spam.