Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key Recall
Recall Date
Jun 10, 2026
Company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Source
FDA
Units
27 units
Reason: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Reason for Recall
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 27 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated June 16, 2026
- Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2288-2026 · 51 units units
- Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2284-2026 · 16 units units
- Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key · ID FDA-DEV-Z-2285-2026 · 27 units units
More recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Philips Azurion 7 B20. Model Numbers: 722068, 722226. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 7 M12. Model Numbers: 722078, 722223. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M12. Model Number: 722227. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M20. Model Numbers: 722228, 722281. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion S. Model Number: 722285. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.