OmniTract Surgical Retractor System, which includes 633 of the firm s product codes. Recall
Recall Date
Jul 15, 2026
Company
Integra LifeSciences Corp. (NeuroSciences)
Source
FDA
Units
633 units
Reason: The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Reason for Recall
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 633 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated July 22, 2026
More recalls from Integra LifeSciences Corp. (NeuroSciences)
- Cytal Burn Matrix 10x15 cm. Product ID: BMM1015. · Integra LifeSciences Corp. (NeuroSciences) · Apr 22, 2026
- Cytal Wound Matrix 2-Layer 5x5 cm. Product ID: WSM0505. · Integra LifeSciences Corp. (NeuroSciences) · Apr 22, 2026
- Cytal Burn Matrix 7x10 cm. Product ID: BMM0710. · Integra LifeSciences Corp. (NeuroSciences) · Apr 22, 2026
Similar recalls
- Philips Azurion 7 M12. Model Numbers: 722078, 722223. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M20. Model Numbers: 722228, 722281. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C · Canon Medical System, USA, INC. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V · Canon Medical System, USA, INC. · Nov 4, 2026
- CareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantable · Medtronic, Inc. · Nov 4, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.