Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders Recall
Recall Date
Jul 22, 2026
Company
Daavlin Distributing Company
Source
FDA
Units
212 units
Reason: Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Reason for Recall
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 212 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated July 27, 2026
- Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders · ID FDA-DEV-Z-2743-2026 · 26 units units
- Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders · ID FDA-DEV-Z-2744-2026 · 212 units units
- Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders · ID FDA-DEV-Z-2742-2026 · 2 units units
More recalls from Daavlin Distributing Company
- Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders · Daavlin Distributing Company · Jul 22, 2026
- Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders · Daavlin Distributing Company · Jul 22, 2026
- Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physici · Daavlin Distributing Company · Jun 10, 2026
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.