Optima XR220amx, Mobile Digital-Ready Radiographic System Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Jul 22, 2026

Company

GE Medical Systems, LLC

Source

FDA

Units

38 units

Reason: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Reason for Recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

Product: Optima XR220amx, Mobile Digital-Ready Radiographic System
Official Recall ID: FDA-DEV-Z-2727-2026
Units affected: 38 units

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 38 units

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated July 27, 2026

All affected products in this recall
  • Optima XR220amx, Mobile Digital-Ready Radiographic System · ID FDA-DEV-Z-2727-2026 · 38 units units
  • Optima XR200amx, Mobile Digital-Ready Radiographic System · ID FDA-DEV-Z-2726-2026 · 4 units units