Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEON Recall
Recall Date
Sep 2, 2026
Company
Medline Industries, LP
Source
FDA
Units
145525 units
Reason: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Reason for Recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 145525 units
5-step guide to help you dispose this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated September 8, 2026
More recalls from Medline Industries, LP
- MEDLINE medical convenience kits labeled as (list truncated) · Medline Industries, LP · Oct 28, 2026
- MEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64155B. · Medline Industries, LP · Oct 28, 2026
- Medline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D · Medline Industries, LP · Oct 21, 2026
- Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135A CHN GYN ONCOLOGY DYNJ906731M · Medline Industries, LP · Oct 21, 2026
- Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D · Medline Industries, LP · Oct 21, 2026
Similar recalls
- Philips Azurion 7 M12. Model Numbers: 722078, 722223. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M20. Model Numbers: 722228, 722281. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C · Canon Medical System, USA, INC. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V · Canon Medical System, USA, INC. · Nov 4, 2026
- CareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantable · Medtronic, Inc. · Nov 4, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.