Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665 Recall
Recall Date
Sep 23, 2026
Company
Datex-Ohmeda, Inc.
Source
FDA
Units
410 units
Reason: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Reason for Recall
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 410 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated September 29, 2026
- Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000 · ID FDA-DEV-Z-3190-2026 · 140 units units
- Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665 · ID FDA-DEV-Z-3192-2026 · 410 units units
- Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 · ID FDA-DEV-Z-3191-2026 · 7 units units
More recalls from Datex-Ohmeda, Inc.
Similar recalls
- Philips Azurion 7 M12. Model Numbers: 722078, 722223. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- Philips Azurion 5 M20. Model Numbers: 722228, 722281. · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C · Canon Medical System, USA, INC. · Nov 4, 2026
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V · Canon Medical System, USA, INC. · Nov 4, 2026
- CareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantable · Medtronic, Inc. · Nov 4, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.