Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665 Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Sep 23, 2026

Company

Datex-Ohmeda, Inc.

Source

FDA

Units

410 units

Reason: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Reason for Recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Product: Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
Official Recall ID: FDA-DEV-Z-3192-2026
Units affected: 410 units

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 410 units

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated September 29, 2026

All affected products in this recall
  • Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000 · ID FDA-DEV-Z-3190-2026 · 140 units units
  • Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665 · ID FDA-DEV-Z-3192-2026 · 410 units units
  • Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 · ID FDA-DEV-Z-3191-2026 · 7 units units

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored

Stay Informed

Weekly recall digest — free, no spam.