Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440 Recall
Recall Date
Oct 28, 2026
Company
Boston Scientific Cork, Ltd.
Source
FDA
Units
620
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Reason for Recall
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.