Techno-Path Multichem AE Unassayed Level 2 REF: AE622X Recall
Recall Date
Oct 28, 2026
Company
TECHNO-PATH MANUFACTURING LTD.
Source
FDA
Units
123 units
Reason: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Reason for Recall
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 123 units
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated November 3, 2026
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.