Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Oct 28, 2026

Company

Medtronic MiniMed, Inc.

Source

FDA

Units

623

Reason: A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

Reason for Recall

A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

What Should I Do?

Quantity: 623

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated November 3, 2026

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored

Stay Informed

Weekly recall digest — free, no spam.