EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Nov 4, 2026

Company

BioMerieux SA

Source

FDA

Units

33

Reason: Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Reason for Recall

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Product: EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
Official Recall ID: FDA-DEV-Z-0301-2027
Units affected: 33

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 33

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated November 10, 2026

Injured by a recalled product?

Our legal partners offer free consultations for recall victims.

Get Free Legal Help →Sponsored

Stay Informed

Weekly recall digest — free, no spam.