VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006 Recall
This is part of a larger recall. See all affected products.
Recall Date
Oct 14, 2026
Company
Steris Corporation
Source
FDA
Units
266
Reason: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Class II: Use of the product may cause temporary or medically reversible health problems.
Official FDA recall noticeInformation is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.
Description
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Reason for Recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
What Should I Do?
Check the model/lot number and contact your healthcare provider or the manufacturer.
Quantity: 266
5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.
+2 more steps…
Last updated October 21, 2026
More recalls from Steris Corporation
- HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18 · Steris Corporation · Oct 14, 2026
- VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060 · Steris Corporation · Oct 14, 2026
- VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007 · Steris Corporation · Oct 14, 2026
- VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061 · Steris Corporation · Oct 14, 2026
- VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 · Steris Corporation · Oct 14, 2026
Similar recalls
- TruSystem 7000 U Operating Table (dV), Product Code 1723633 · Baxter Healthcare Corporation · Nov 11, 2026
- Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, · Cochlear Limited · Nov 11, 2026
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8698P (HSI Item No. 600-5659); Product Description: Sterile, synthetic, nonabs · Henry Schein Inc. · Nov 11, 2026
- Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: SWAN GANZ, KIT, CATALOG # PC37SGPHA · Cardinal Health 200, LLC · Nov 11, 2026
- CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010 · Artivion, Inc · Nov 11, 2026
Injured by a recalled product?
Our legal partners offer free consultations for recall victims.
Stay Informed
Weekly recall digest — free, no spam.