VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061 Recall

Ongoing
Medical Devices
Class II — Moderate

Recall Date

Oct 14, 2026

Company

Steris Corporation

Source

FDA

Units

2205

Reason: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Class II: Use of the product may cause temporary or medically reversible health problems.

Official FDA recall notice

Information is from official FDA notices. Do not stop taking any medication without consulting your doctor or pharmacist first — the risks of stopping may be greater than the risk from the recalled product.

Recall Details

Description

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Reason for Recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Product: VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
Official Recall ID: FDA-DEV-Z-0073-2027
Units affected: 2205

What Should I Do?

Check the model/lot number and contact your healthcare provider or the manufacturer.

Quantity: 2205

Remediation GuideAction Required

5-step guide to help you action required this recalled product. Click Start Remediation to track your progress.

1
Confirm Your Product Is Affected
2
Read the Full Recall Notice
3
Stop Using the Product if in Doubt

+2 more steps…

View Official Recall Notice (FDA)

Last updated October 21, 2026

All affected products in this recall
  • VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 · ID FDA-DEV-Z-0074-2027 · 217 units
  • VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006 · ID FDA-DEV-Z-0070-2027 · 266 units
  • VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061 · ID FDA-DEV-Z-0073-2027 · 2205 units
  • VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007 · ID FDA-DEV-Z-0071-2027 · 409 units
  • VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060 · ID FDA-DEV-Z-0072-2027 · 1185 units